Instructions for use

How to use and replace Yewtwist.

Disconnecting #1

Description

A device for disconnecting IV, dialysis and feeding connections. Provides users extra grip and leverage, so connections come apart with less strain on users hands and wrists, and lowers the risk of damaging IV lines.

  1. Straight bar design gives added strength.
  2. Soft, non-slip jaws provide a secure grip on the connection without damaging the line.
  3. Three jaw sizes fit a wide range of IV, dialysis and feeding tube connections.

Indications for use

To limit force on the hand and wrist, use Yewtwist on all IV disconnections.

  • For routine use on every disconnection, and on overtightened, difficult to disconnect or adhered fittings.
  • Intended to assist with the disconnection of Luer lock fittings, including fittings used for medical purposes, by providing additional grip and leverage.

Approved use

Single patient use only

For the highest level of aseptic practice, dedicate one Yewtwist to each patient for the full course of their therapy.

  • Keep it at the bedside or with the patient’s own supplies. Do not keep it in a shared drawer, cart or pocket.
  • Write the patient’s name on the label of the resealable bag.
  • Clean before and after each use with hospital-grade wipes.
  • Discard when the patient’s therapy ends, or replace sooner if worn.
  • Do not share between patients.

Directions for use

  1. Gloved hands wiping a green Yewtwist with a disinfectant wipe

    Clean Yewtwist

    Wipe Yewtwist with a hospital-grade disinfectant wipe.

  2. Gloved hands holding Yewtwist with the small, medium and large jaws labelled

    Select jaw size

    Choose the jaw that best fits the connector. Place it on the connector furthest from the patient.

  3. Gloved hands pinching Yewtwist on a connector, with an arrow showing the counterclockwise turn

    Pinch and turn

    Pinch the arms together and turn counterclockwise.

  4. Gloved hands separating the two halves of a connection after using Yewtwist

    Detach

    Separate the connection.

Cleaning instructions

Clean Yewtwist before and after each use with a hospital-grade disinfectant wipe. Most hospital-grade disinfectant wipes may be used.

  1. Wipe all surfaces thoroughly, including the jaws and the grip ridges.
  2. Follow the contact time in the wipe manufacturer’s instructions for use.
  3. Allow to air dry before use or storage.

Tested disinfectants

  • Diversey™ Oxivir® Tb RTU Disinfectant Cleaner
  • Clorox Healthcare® Bleach Germicidal Cleaner
  • Clorox Healthcare® Hydrogen Peroxide Cleaner
  • Super Sani-Cloth® Wipes
  • Metrex CaviWipes™

Inspection and replacement

Inspect Yewtwist before each use. Replace it if the jaws are cracked, torn, no longer grip, or if the body is cracked or bent.

Warnings, precautions and storage

  • Nonsterile.
  • Not made with natural rubber latex.
  • Pinch point hazard.
  • Do not apply excessive force when attempting to loosen a connection that is firmly seized or cannot be safely disconnected.
  • Do not pull on the line or tube while Yewtwist is attached.
  • Do not use Yewtwist to tighten connections. Connect by hand.
  • Not a toy. Keep away from small children and pets.
  • For hospital, clinic and home use.
  • Store clean and dry at room temperature, in its bag.